6.0 Sample Test Procedures
Before performing the sample tests, please be sure to read Appendix E. It contains instructions on installing the GUI and config-
uring the hardware completely. Note: The on-board FPGAs and PSoC will be preprogrammed at the factory for testing. Therefore
there should be no need to program them. Appendix E contains programming instructions should you need them.
6.1 Required Equipment
• Tektronix Waveform Monitor: WFM 700
• Quad Independent Channel HOTLink II CYV15G0404DXB Video PHY Demo Board
• PC/Laptop with HVDB installed
• USB cable
• 6V DC Power Supply
• Four–six BNC cables
6.2 Tektronix WFM 700
The Tektronix WFM 700 is used in the demo tests to display the signal(s) output by the video demo board on the channel(s)
specified by the user. As mentioned in previous sections, the WFM 700 does not support data rates of 360 Mb/s and 540 Mb/s.
For a brief description of relevant buttons for the tests, please see Figure 6-1. The connectors used in the tests are shown in
Figure 6-2.